GAMMA-SCOUT® geiger counter dosimeter
Item number: I 624
Price: EUR 350.00
GAMMA-SCOUT meaures and logs radioactivity ongoing. Results can be displayed as micro-Sievert, or as counts. By means of a aperture system the separate measurement of alpha, beta and gamma radiation is possible.
Basic version WITHOUT ticker and alert
Large range of calibration
- In measuring mode, the display shows the current radiation measurement reading. The display also temporarily shows the average H of the last day (24h, 12 am to 12 am).
- GAMMA-SCOUT® is calibrated across a wide scale (0.01 up to1000.00 µSv/h).
- Each GAMMA-SCOUT® radiation meter is subjected to a final test. This test is supervised by the German Institute of Radiation Protection; a government controlled university for Applied Technology. The tested device must be in a confidence interval of 5% in comparison to a master. This master is adjusted to a gauged reference Cs-137 emitter.
- The conversion from impulses per time into the dose rate is based on Cs-137. If other isotopes are present, the impulses per time can be directly displayed, and can be converted by using personal charts.
- GAMMA-SCOUT® can be used as a dose meter (cumulative radiation).
- The dose rate may be displayed in Sievert or Rem.
- The GAMMA-SCOUT® monitors radiation day and night and logs data for later download. Due to sophisticated electronics, its battery lasts for years.
- GAMMA-SCOUT® stores all registered pulses in its internal memory and keeps them ready for you to use when desired. 32,000 Readings: For constant monitoring of the environmental data, the interval of data storage is set on a value of one per week. This equals 600 years of memory capacity. The user can set up smaller intervals, which maxes out the memory capacity sooner. To clear the memory content, see the manual.
- Liquid-crystal display (LCD), 4-digit, numeric with dimension,
- quasi-analogue logarithmic bar chart.
- Operating mode indicators
- End-window alpha-beta-gamma detector counting tube according to the Geiger-Müller principle
- Stainless steel housing
- Measuring length 38.1 mm, measuring diameter 9.1 mm
- Mica window 1.5 to 2 mg/cm2
- Zero rate <10 pulses per minute with sreening by 3mm AI and 50mm Pb Operating temperature -20 bis +60°C, operating voltage approx. 450 V Calibrated scale 0.01 μSv/h up to 1000.00 μSv/h
- alpha: from 4 MeV
- beta: from 0.2 MeV
- gamma: from 30 keV
- alpha+beta+gamma: without shielding
- beta+gamma: Al foil approx. 1mm, shields off completely
- gamma: Al shielding approx. 3 mm, Shielding off a completely and ß to 2 MeV, weakens y less than 7% based on Cs-137
- Less than 10 microamperes under environmental radiation
- 64 Kbyte (32,000 data sets)
- impact-resistant plastic
Weight | 425 g. |
---|